Memo-Graph S 10 Technical data
Endress+Hauser 87
10.3 Output values
10.3.1 Multifunction output board (socket 2)
Analogue outputs 4 or 8 analogue outputs, each galvanically isolated from all electric circuits (test voltage 500 V
AC
)
Output range: 0 - 10 V, 1 - 5 V, 0 - 20 mA, 4 - 20 mA
Accuracy: 0.25 % of range (at 1-5 V 0.5 %)
Resolution: 0.025 % (at 1-5 V 0.06 %)
Temperature drift: < 0.05 % /K of range
Output ripple: < 10 mV effective
Response time: max. 300 ms (leap at input 10 % -> 90 % of measuring range)
Load resistance (current output): max. 500 Ohm
Output current (voltage output): min. 10 mA
Loop power supply Software selectable for each channel (alternative to analogue output)
Number of transmitter power supplies: 4 or 8, each galvanically isolated from all circuits
Output voltage: 24 V
DC
±15 %, open loop <28 V;
Output current: max. 25 mA (internal current limiter), short-circuit proof
Relay outputs Number of relay outputs: 6 relays, NO contact (230 V / 3 A, insulation group A as per VDE 0110)
mixture of SELV circuits and mains circuits not permitted. Can be configured as NC contacts via
operation.
10.3.2 Digital I/O board (socket 1, socket 2)
Relay outputs Number of relay outputs: 6 relays, NO contact, 230 V / 3 A, for limit value condition. Mixture of
SELV circuits and mains circuits not permitted. Can be configured as NC contacts via operation.
10.3.3 Power supply board (socket 3)
Auxiliary voltage output for digital input control with potential-free contacts,
24 V
DC
±15 %, max. 100 mA, open loop <28 V, short-circuit proof, unstabilised
Relay outputs 4 relays, NO contact, 230 V / 3 A, for limit value condition. Mixture of SELV circuits and mains
circuits not permitted. Can be configured as NC contact via operation
1 open collector output (max. 100 mA / 25 V)
Common relay 1 relay, changeover contact 230 V / 3 A, for limit value condition / power failure
10.4 Power supply
Electrical connection
(circuit diagram)
See “Connections / terminal diagram” in Chapter 4.1
Supply voltage / Power
consumption
Low voltage power unit: 115 to 230 V
AC
(+10% -15%), 50/60 Hz, max. 25 VA (full version)
Extra-low voltage power unit: 24 V
AC/DC
(+20% -15%), 0/50/60 Hz, max. 25 VA (full version)
Electrical safety EN 61010-1, protection class I, overvoltage category II
Cable specifications /
Connections
Keyed screw plug-in terminal strips,
Wire cross section on analogue inputs / digital I/O max. 1.5 mm² / 0.0023 in²,
Power supply / relays max. 2.5 mm² / 0.0039 in² (each with ferrules)
Interface connections
Front-mounted RS 232 interface (3.5 mm / 0.14“ stereo jack plug, only for IP 54 front with door)
Rear-mounted RS 232 interface (9 pin, Sub-D, socket)
Serial interface (option) RS 485 (rear-mounted) adjustable device address;
Cable length max. 1000 m / 0.62 mi screened cable
Ethernet (option) Internal Ethernet interface, 10BaseT, plug type RJ45